Small field of view detectors based on Position Sensitive Photomultiplier Tubes (PSPMTs) have been widely used by many research groups in small-animal imaging and recently clinically in scintimammography. In most cases the PSPMT is coupled to pixelized scintillators since it has been shown that their use improves spatial resolution. The use of a single energy window for each crystal element is the only proposed method for scatter correction in pixelized scintillators. In this work we have modified the dual energy window subtraction technique in order to be applicable to a pixelized scintillator detector and we have evaluated its performance with real data. Hot, cold and breast phantoms have been imaged in planar mode.
机译:基于位置敏感光电倍增管(PSPMT)的小视场检测器已被许多研究小组广泛地用于小动物成像以及最近在闪烁X线照相术中的临床应用。在大多数情况下,PSPMT与像素化闪烁体耦合,因为已经证明它们的使用可以提高空间分辨率。对每个晶体元素使用单个能量窗口是像素化闪烁体中散射校正的唯一建议方法。在这项工作中,我们修改了双能量窗口减法技术以适用于像素化闪烁体检测器,并使用实际数据评估了其性能。热,冷和乳房幻影已在平面模式下成像。