首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >High-speed single photon counting read out electronics for a digital detection system for clinical synchrotron radiation mammography
【24h】

High-speed single photon counting read out electronics for a digital detection system for clinical synchrotron radiation mammography

机译:高速单光子计数读出电子设备,用于临床同步辐射乳腺摄影的数字检测系统

获取原文
获取原文并翻译 | 示例

摘要

The SYRMEP beam line is currently in the upgrading phase for mammographic examinations on patients at Elettra in Trieste. At the same time, a digital detection system, suitable for in -vivo breast imaging, is under development; it consists of a silicon laminar detector array operating in single photon counting mode. The duration of a clinical examination should not exceed a few seconds. Fast read out electronics is therefore necessary with the aim of avoiding losses in image contrast in presence of high counting rates. A custom ASIC working with 100% efficiency for rates up to 100 kHz per pixel has been designed and tested, and other solutions based on commercially available ASICs are currently under test. Several detector prototypes have been assembled, and images of mammographic test objects have been acquired. Image quality, efficiency and contrast losses have been evaluated in all cases as a function of the counting rate.
机译:SYRMEP光束线目前处于升级阶段,以进行对的里雅斯特Elettra患者的乳房X线检查。同时,正在开发一种适用于体内乳房成像的数字检测系统。它由以单光子计数模式运行的硅层状探测器阵列组成。临床检查的持续时间不应超过几秒钟。因此,为了避免在高计数率的情况下图像对比度的损失,必须快速读取电子电路。已经设计和测试了一种定制ASIC,该ASIC以100%的效率工作,每个像素的速率高达100 kHz,目前正在测试基于商用ASIC的其他解决方案。已经组装了几个探测器原型,并已获取了乳房X光检查对象的图像。在所有情况下,都根据计数率评估了图像质量,效率和对比度损失。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号