Photodiodes array as potential diagnostic for measuring short bursts of K-α radiation from Ti targets irradiated with 45 fs laser pulses

机译:光电二极管阵列作为潜在的诊断工具,可测量来自45 fs激光脉冲照射的Ti目标的K-α短脉冲

获取原文
获取原文并翻译 | 示例

摘要

We report on a study comparing absolute K-a yield from Ti foils measured with a calibrated system of an X-ray CCD coupled to a curved LiF Von-Hamos crystal spectrometer to the difference in the signals measured simultaneously with two similar photodiodes fitted with two different filters. Our data indicate that a combination of photodiodes with different filters could be developed into an alternative and inexpensive diagnostic for monitoring single shot pulsed emission in a narrow band of X-ray region.
机译:我们报告了一项研究,该研究比较了使用校准的X射线CCD校准系统和弯曲的LiF Von-Hamos晶体光谱仪耦合测得的Ti箔片的绝对Ka产率,以及使用两个类似的装有两个不同滤光片的光电二极管同时测量的信号差。我们的数据表明,光电二极管与不同滤光片的组合可以发展成为一种替代且廉价的诊断工具,用于监测X射线区域窄带中的单脉冲发射。