Experimental study of beam hardening artifacts in photon counting breast computed tomography

机译:束硬化假象在光子计数乳房计算机断层扫描中的实验研究

获取原文
获取原文并翻译 | 示例

摘要

We are implementing an X-ray breast Computed Tomography (CT) system on the gantry of a dedicated single photon emission tomography system for breast Tc-99 imaging. For the breast CT system we investigated the relevance of the beam hardening artifact. We studied the use of a single photon counting silicon pixel detector (0.3 mm thick, 256 × 256 pixel, 55μm pitch, bump-bonded to the Medipix2 photon counting readout chip) as detector unit in our X-ray CT system. We evaluated the beam hardening "cupping" artifact using homogeneous PMMA slabs and phantoms up to 14 cm in diameter, used as uncompressed breast tissue phantoms, imaged with a tungsten anode tube at 80 kVp with 4.2 mm Al filtration. For beam hardening evaluation we used a bimodal energy model. The CT data show a "cupping" artifact going from 4% (4-cm thick material) to 18% (14-cm thick material). This huge artifacts is influenced by the low detection efficiency and the charge sharing effect of the silicon pixel detector.
机译:我们正在为专用于乳房Tc-99成像的专用单光子发射断层扫描系统的机架上实施X射线乳房计算机断层扫描(CT)系统。对于乳房CT系统,我们研究了束硬化伪影的相关性。我们研究了在X射线CT系统中使用单个光子计数硅像素检测器(0.3 mm厚,256×256像素,55μm间距,凸点结合到Medipix2光子计数读出芯片)作为检测器单元的情况。我们使用均匀的PMMA平板和直径最大14厘米的幻像(用作未压缩的乳腺组织幻像)评估了束硬化的“杯状”伪影,这些幻像用80 kVp的钨阳极管和4.2毫米Al过滤成像。对于束硬化评估,我们使用了双峰能量模型。 CT数据显示“杯形”伪影从4%(4厘米厚的材料)到18%(14-cm厚的材料)。这种巨大的伪像受到硅像素检测器的低检测效率和电荷共享效应的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号