Efficiency calibration of an extended-range Ge detector by a detailed Monte Carlo simulation

机译:通过详细的蒙特卡洛模拟对大范围锗探测器的效率校准

获取原文
获取原文并翻译 | 示例
           

摘要

A Monte Carlo simulation has been employed for calibrating an extended-range Ge detector in an energy range from 14 to 1800keV. A set of sources from monoenergetic and multi-gamma emitters point were measured at 15 cm from the detector window and provided 26 experimental values to which the results of the simulations are compared. Discrepancies between simulated and experimental values are within 1 standard deviation, and relative differences are, in most cases, below 1%.
机译:蒙特卡罗模拟已用于校准14至1800keV能量范围内的扩展范围的Ge检测器。在距探测器窗口15 cm处测量了一组来自单能和多伽玛发射源的光源,并提供了26个实验值,将其与模拟结果进行了比较。模拟值与实验值之间的差异在1个标准偏差内,并且在大多数情况下,相对差异在1%以下。