首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >Evaluation of image performance of CZT detector for digital mammography: Monte Carlo simulation
【24h】

Evaluation of image performance of CZT detector for digital mammography: Monte Carlo simulation

机译:数字乳腺摄影CZT检测器的图像性能评估:蒙特卡洛模拟

获取原文
获取原文并翻译 | 示例

摘要

Hybrid pixel detector arrays that convert X-ray photons directly into electrical signal are still at the developmental stage for diagnostic X-ray applications. In this work, the feasibility of using CZT detector for application to digital mammography is investigated. In order to evaluate the proposed detection system, a simulation of X-ray photon transport in the sensor material was conducted using the extended version of Monte Carlo N-Particle (MCNPX) code. Charge carrier transport was modelled through the solution of Poisson's equation. The two models were then combined in order to evaluate the response of the detector. The line spread function (LSF) was calculated in order to estimate the modulation transfer function (MTF) of the system. CdTe and Si detectors are included for comparison and two different pixel pitches are studied.
机译:将X射线光子直接转换为电信号的混合像素检测器阵列仍处于诊断X射线应用程序的开发阶段。在这项工作中,研究了将CZT检测器应用于数字乳腺摄影的可行性。为了评估建议的检测系统,使用蒙特卡罗N粒子(MCNPX)代码的扩展版本对传感器材料中X射线光子传输进行了模拟。通过泊松方程的求解对载流子传输进行建模。然后将两个模型结合起来,以评估检测器的响应。计算线路扩展函数(LSF),以估计系统的调制传递函数(MTF)。包括CdTe和Si检测器进行比较,并研究了两个不同的像素间距。