首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators
【24h】

An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators

机译:一种精确校准放射治疗直线加速器机架角度指示器的改进方法

获取原文
获取原文并翻译 | 示例
           

摘要

The calibration of the gantry angle indicator is an important and basic quality assurance (QA) item for the radiotherapy linear accelerator. In this study, we propose a new and practical method, which uses only the digital level, V-film, and general solid phantoms. By taking the star shot only, we can accurately calculate the true gantry angle according to the geometry of the film setup. The results on our machine showed that the gantry angle was shifted by -0.11° compared with the digital indicator, and the standard deviation was within 0.05°. This method can also be used for the simulator. In conclusion, this proposed method could be adopted as an annual QA item for mechanical QA of the accelerator.
机译:机架角度指示器的校准是放射治疗线性加速器的重要且基本的质量保证(QA)项目。在这项研究中,我们提出了一种新的实用方法,该方法仅使用数字水准仪,V胶片和一般的实物模型。通过仅拍摄星形镜头,我们可以根据胶片设置的几何形状准确计算出真正的龙门角度。我们机器上的结果表明,与数字指示器相比,机架角度偏移了-0.11°,标准偏差在0.05°以内。此方法也可以用于模拟器。总之,该方法可以作为加速器机械质量保证的年度质量保证项目。