A solid-state Cadmium Zinc Telluride (CZT) gamma camera in conjunction with a Modified Uniformly Redundant Arrays (MURAs) Coded Aperture (CA) scintimammography (SM) system has been investigated using Monte Carlo simulation. The motivation is to utilise the enhanced energy resolution of CZT detectors compared to standard scintillation-based gamma cameras for scatter rejection. The effects of variations in lesion sizes and tumour-to-background-ratio were simulated in a 3D phantom geometry. Despite the enhanced energy resolution, we find that the open field geometry associated with the MURA CA imaging nonetheless requires shielding from nonspecific background tracer uptake, and correction for out-of-plane activity. We find that a TBR of 20:1 is required for visualising a 10 mm wide lesion.
机译:使用蒙特卡洛模拟研究了固态碲化锌锌(CZT)伽马相机,结合改进的均匀冗余阵列(MURAs)编码孔径(CA)闪烁X线照相(SM)系统。与基于标准闪烁的伽马相机相比,其动机是利用CZT检测器提高的能量分辨率来进行散射抑制。在3D幻像几何体中模拟了病变大小和肿瘤与背景的比率变化的影响。尽管提高了能量分辨率,但我们发现与MURA CA成像相关联的开放视野几何形状仍需要屏蔽非特异性背景示踪剂的吸收以及对平面外活动的校正。我们发现可视化10毫米宽的病变需要20:1的TBR。