Performance of a compact injector for heavy-ion medical accelerators

机译:用于重离子医用加速器的紧凑型注射器的性能

获取原文
获取原文并翻译 | 示例
           

摘要

A compact injector, designed for a heavy-ion medical accelerator complex, was constructed. It consists of an Electron-Cyclotron-Resonance Ion-Source (ECRIS) and two linacs, which are a Radio-Frequency-Quadrupole linac and an Interdigital H-mode Drift-Tube-Linac (IH-DTL) having the same operating frequency of 200 MHz. For beam focusing of the IH-DTL, the method of Alternating-Phase-Focusing (APF) was employed. The compact injector can accelerate heavy ions having a charge-to-mass ratio of q/m = 1/3 up to 4.0 MeV/u. Use of the APF IH-DTL and operating frequency of 200 MHz allowed us to design compact linacs; the total length of the two linacs is less than 6 m. Beam-acceleration tests of the compact injector system were performed. The measured intensity of accelerated ~(12)C~(4+) beams with the compact injector was 380 eμA. Beam transmission of the APF IH-DTL was estimated to be as high as 96%, which is comparable to the value calculated by a simulation code. Transverse phase-space and energy distributions of accelerated beams were measured and compared with those calculated by the simulation code, and we found that they were agreed well with each other.
机译:构造了一种专为重离子医用加速器综合体设计的紧凑型注射器。它由电子回旋共振离子源(ECRIS)和两个直线加速器组成,分别是射频四极直线加速器和叉指H模式漂移管直线加速器(IH-DTL),其工作频率与200 MHz。对于IH-DTL的光束聚焦,采用了交替相聚焦(APF)的方法。紧凑型注射器可以加速质荷比为q / m = 1/3的重离子加速至4.0 MeV / u。使用APF IH-DTL和200 MHz的工作频率使我们能够设计紧凑的直线加速器;两个直线加速器的总长度小于6 m。进行了紧凑型喷射器系统的光束加速测试。紧凑型注射器测量的〜(12)C〜(4+)加速光束强度为380eμA。 APF IH-DTL的光束透射率估计高达96%,与模拟代码计算出的值相当。测量了加速光束的横向相空间和能量分布,并将其与仿真代码计算的结果进行比较,我们发现它们彼此吻合良好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号