Monolithic scintillator blocks provide an efficient way to significantly increase the sensitivity of high-resolution positron emission tomography (PET) systems. Using statistical or neural network-based positioning schemes, the measured intrinsic detector resolution for perpendicular incident photons is 1.7mm full-width at half-maximum (FWHM) for 10mm thick LSO blocks and 2.0mm FWHM for 20 mm thick LSO blocks. Because the positioning algorithms determine the incidence position as opposed to the combination of interaction position and interaction depth, a very good parallax correction can be achieved. Energy and detector time resolution were 11.5% FWHM and 1.6 ns FWHM, respectively.
机译:单片闪烁器模块提供了一种有效的方法,可以显着提高高分辨率正电子发射断层扫描(PET)系统的灵敏度。使用基于统计或神经网络的定位方案,对于10mm厚的LSO块,垂直入射光子的实测本征检测器分辨率为半峰全宽(FWHM)1.7mm,对于20mm厚的LSO块为2.0mm FWHM。因为定位算法确定入射位置,而不是交互位置和交互深度的组合,所以可以实现非常好的视差校正。能量和检测器时间分辨率分别为11.5%FWHM和1.6 ns FWHM。