Two-dimensional multiwire gas proportional detector for X-ray photon correlation spectroscopy of condensed matter

机译:X射线光子相关光谱的二维多线气体比例探测器

获取原文
获取原文并翻译 | 示例

摘要

Details of a two-dimensional (2-D) multiwire gas proportional detector for X-ray photon correlation spectroscopy (XPCS) of condensed matter are described. The characteristics of the gas proportional detector at 8 keV, 0.3 pC anode charge, and 3 bar (absolute) of Xe/10%CO_2 are as follows: 8.5 × 10~(-7)counts/s (100 × 100 μm~2) dark count rate, ~μs time resolution, ~48 and 73 μm position resolution (FWHM) along and across the anode wire direction, respectively, and ~80% quantum efficiency. The effects of incident photon energy, anode charge (i.e., gain), gas drift depth, and gas pressure on position resolution are discussed. Static and dynamic speckle patterns, measured from disordered aerogel and polystyrene/polybutadiene blends by a partially coherent synchrotron X-ray source, demonstrate that a 2-D multiwire gas proportional detector is very suitable for the dynamic study of condensed matter with relaxation times in the order of μs to 10~3s and atomic length scale.
机译:描述了用于冷凝物的X射线光子相关光谱(XPCS)的二维(2-D)多线气体比例探测器。气体比例检测器在8 keV,阳极电荷0.3 pC和3 bar(绝对)Xe / 10%CO_2时的特性如下:8.5×10〜(-7)counts / s(100×100μm〜2 )暗计数率,〜μs时间分辨率,分别沿阳极线方向和跨阳极线方向的〜48和73μm位置分辨率(FWHM),以及〜80%的量子效率。讨论了入射光子能量,阳极电荷(即增益),气体漂移深度和气体压力对位置分辨率的影响。通过部分相干的同步加速器X射线源从无序的气凝胶和聚苯乙烯/聚丁二烯混合物中测得的静态和动态散斑图表明,二维多线气体比例检测器非常适合于动态研究冷凝物的弛豫时间。 μs到10〜3s的量级和原子长度标度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号