摘要

The polarization analysis options on the thermal neutron triple-axis spectrometer TAS-1 at JRR-3 of Japan Atomic Energy Agency (JAEA) have been improved as a result of an upgrading program of the Advanced Science Research Center. In order to increase the polarized neutron intensity, we have replaced the previous flat polarizing Cu_2MnAl (Heusler) monodhromator and analyzer with doubly focusing ones. Intensity gain factors of more than 30 have been achieved for 14.7 meV neutrons in measurements of incoherent neutron scattering of vanadium. Using the new monochromator and analyzer, elastic and inelastic polarized neutron scattering experiments with the conventional setup for one-dimensional polarization analysis and with CRYOPAD (Cryogenic Polarization Analysis Device) for three-dimensional polarization analysis are now routinely performed on TAS-1.
机译:由于高级科学研究中心的升级计划,日本原子能机构(JAEA)JRR-3的热中子三轴光谱仪TAS-1的极化分析选项得到了改进。为了增加极化中子的强度,我们用双重聚焦装置代替了以前的平面极化Cu_2MnAl(Heusler)单偏光片和分析仪。在测量钒的非相干中子散射时,对于14.7 meV中子,已经获得了超过30的强度增益因子。使用新的单色仪和分析仪,现在可以在TAS-1上常规进行具有一维极化分析的常规设置和具有CRYOPAD(低温极化分析设备)的弹性和非弹性极化中子散射实验。