首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >First qualification and selection of the eROSITA PNCCDs
【24h】

First qualification and selection of the eROSITA PNCCDs

机译:eROSITA PNCCD的首次鉴定和选择

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

For the X-ray astronomy instrument eROSITA a framestore PNCCD was developed by the MPI Halbleiterlabor. The PNCCD has an image area of 384 × 384 pixels with a size of 75 μm × 75 μm. Each channel of the PNCCD has an own readout anode which allows parallel amplification and signal processing of the CCD signals of one row. The first measurements for the spectroscopic characterization of the PNCCDs are made with a special measurement setup—the so-called Cold Chuck Probe Station. The Cold Chuck Probe Station allows to fully operate the CCD without mounting and bonding the chip on a PCB as the CCD is contacted only with needles. Thus all eROSITA PNCCDs can be qualified under the same measurement conditions and with an identical electronic setup. Therefore the results can be compared directly. The spectroscopic properties of the PNCCDs, like the charge transfer efficiency and the energy resolution are measured. Also pixel defects such as bright pixels or non-transferring pixels are detected. With the Cold Chuck Probe Station a readout noise of 2.7 e~- ENC can be achieved and reliable measurement results obtained. Based on these results the best PNCCDs will be selected for eROSITA.
机译:对于X射线天文仪器eROSITA,MPI Halbleiterlabor开发了一个框架存储PNCCD。 PNCCD的图像面积为384×384像素,尺寸为75μm×75μm。 PNCCD的每个通道都有一个自己的读出阳极,可以对一行的CCD信号进行并行放大和信号处理。 PNCCD的光谱表征的首次测量是通过特殊的测量设置进行的,即所谓的冷卡盘探针站。由于CCD仅与针头接触,因此冷夹头探针台无需安装芯片并在PCB上粘结即可完全操作CCD。因此,所有eROSITA PNCCD都可以在相同的测量条件和相同的电子设置下进行鉴定。因此,可以直接比较结果。测量了PNCCD的光谱特性,例如电荷转移效率和能量分辨率。另外,还检测到诸如亮像素或非传输像素之类的像素缺陷。使用冷卡盘探针台,可实现2.7 e〜-ENC的读出噪声,并获得可靠的测量结果。基于这些结果,将为eROSITA选择最佳的PNCCD。

著录项

  • 来源
  • 作者单位

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

    MPI Halbkiterlabor. Otto-Hahn-Ring 6, 81739 Muenchen, Germany,Max-Planck-Institut fuer extraterrestrische Physik. Giessenbachstrasse, 85748 Garching, Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    PNCCD; eROSITA; X-ray detector; probe measurement;

    机译:PNCCD;eROSITA;X射线探测器;探针测量;