首页> 外文期刊>Nuclear instruments and methods in physics research >Spatial resolution of the Inveon small-animal PET scanner for the entire field of view
【24h】

Spatial resolution of the Inveon small-animal PET scanner for the entire field of view

机译:Inveon小动物PET扫描仪在整个视场中的空间分辨率

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

In high throughput small-animal positron emission tomography (PET) studies, it can be advantageous to scan more than one animal at the same time. To optimally position these animals in the scanner, both sensitivity and spatial resolution have to be known across the entire field of view (FOV). The Siemens Inveon small-animal PET scanner has been characterized according to NEMA NU 4-2008 standards, but the spatial resolution has only been determined in a small portion of the FOV and only for filtered backprojection (FBP). In this study, the spatial resolution in the entire FOV was determined for the Inveon using FBP, ordered subset expectation maximization in 2 dimensions (OSEM2D) and maximum a posteriori (MAP) reconstructions. OSEM2D produced the best spatial resolution in the axial center of the FOV, but MAP outperformed the two other reconstruction algorithms in terms of uniformity of the resolution.
机译:在高通量小动物正电子发射断层扫描(PET)研究中,同时扫描多于一只动物可能是有利的。为了在扫描仪中最佳定位这些动物,必须在整个视场(FOV)中都知道灵敏度和空间分辨率。西门子Inveon小动物PET扫描仪已根据NEMA NU 4-2008标准进行了表征,但是空间分辨率仅在FOV的一小部分中确定,并且仅用于滤波反投影(FBP)。在这项研究中,使用FBP,二维有序子集期望最大化(OSEM2D)和最大后验(MAP)重构确定Inveon整个FOV的空间分辨率。 OSEM2D在FOV的轴心位置产生了最佳的空间分辨率,但就分辨率的均匀性而言,MAP优于其他两种重构算法。

著录项

  • 来源
  • 作者单位

    Department of Nuclear Medicine, Radboud University Nijmegen Medical Centre, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands Institute of Technical Medicine, University of Twente, P.O. Box 217, 7500 AE Enschede, The Netherlands;

    Department of Nuclear Medicine, Radboud University Nijmegen Medical Centre, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands;

    Department of Nuclear Medicine, Radboud University Nijmegen Medical Centre, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands;

    Institute of Technical Medicine, University of Twente, P.O. Box 217, 7500 AE Enschede, The Netherlands;

    Department of Nuclear Medicine, Radboud University Nijmegen Medical Centre, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    positron emission tomography; small-animal; inveon; spatial resolution; maximum a posteriori reconstruction;

    机译:正电子发射断层扫描小动物内脏空间分辨率最大后验重建;