首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >Investigation of the influence of an inner gas bubble on the response of a liquid scintillation detector to γ-rays and neutrons
【24h】

Investigation of the influence of an inner gas bubble on the response of a liquid scintillation detector to γ-rays and neutrons

机译:研究内部气泡对液体闪烁探测器对γ射线和中子的响应的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The influence of an inert gas bubble inside EJ301 cylindrical cell on the response of the liquid scintillation detector to γ-rays and neutrons has been investigated. It is found experimentally that the response of the detector to radiation varies with the detector placement styles. The explanation proposed in this article is that this variation arises from the change of the light transmission process induced by the motion of the inner gas bubble. This assertion has been supported by the simulations of both radiation energy deposition and light transmission process inside the detector with the GEANT4 code. For the recommended detector placement style when the detector is placed horizontally, the variation of the response is smaller than 1.56%, which will increase to about 22% when the detector is placed facing down and the bubble lies between the scintillator and photomultiplier tube.
机译:研究了EJ301圆柱形电池内部惰性气泡对液体闪烁探测器对γ射线和中子响应的影响。实验发现,检测器对辐射的响应随检测器放置方式而变化。本文提出的解释是,这种变化是由内部气泡运动引起的光传输过程的变化引起的。通过使用GEANT4代码对探测器内部的辐射能沉积和光传输过程进行仿真,从而支持了这一主张。对于推荐的检测器放置方式,当检测器水平放置时,响应的变化小于1.56%,当检测器面向下放置且气泡位于闪烁体和光电倍增管之间时,响应的变化将增加到约22%。