首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >Influence of the active area size and read-out method on the timing performance of SiPMs coupled to LYSO scintillators
【24h】

Influence of the active area size and read-out method on the timing performance of SiPMs coupled to LYSO scintillators

机译:有源区大小和读出方法对与LYSO闪烁体耦合的SiPMs定时性能的影响

获取原文
获取原文并翻译 | 示例
           

摘要

In this work we report on the coincidence resolving time performance of SiPMs with different sizes, produced at FBK, coupled to the same LYSO scintillators. The measurements are performed both with and without the differential leading edge discriminator at three different temperatures, 20 ℃, 0 ℃ and - 20 ℃ The photo-detectors feature an active area of 2 × 2 mm~2 and 4×4 mm~2. The scintillators have a cross-section of 1.8 × 1.8 mm~2 and height of 10 mm. The measurements show that, once we eliminate the effect of noise on the timing measurements, we obtain similar coincidence resolving times for the two SiPM sizes considered. This means that the SiPM capacitance, at least up to 4 × 4 mm~2, is not a limiting factor.
机译:在这项工作中,我们报告了在FBK与相同LYSO闪烁体耦合的不同尺寸的SiPM的重合分辨时间性能。在有和没有差分前缘鉴别器的情况下,在20℃,0℃和-20℃这三种不同温度下进行测量。光电探测器的有效面积为2×2 mm〜2和4×4 mm〜2。闪烁体的横截面为1.8×1.8 mm〜2,高度为10 mm。测量表明,一旦消除了噪声对时序测量的影响,对于所考虑的两种SiPM尺寸,我们将获得相似的重合解决时间。这意味着至少达到4×4 mm〜2的SiPM电容不是限制因素。