Preliminary results concerning the study of timing and energy resolution properties of recent Silicon Photomultipliers (SiPMs) devices by STMicroelectronics are here discussed in view of their possible applications in TOF PET modules and PET/MRI hybrid systems. A pair of devices with 60 × 60 microcells and an active area of 3.5 × 3.5 mm~2 have been coupled to 3×3×15 mm~3 LYSO crystals and fully characterized. Measurements of the linearity response with different gamma sources and of the timing resolution with the two 511 keV photons from a ~(22)Na positron source have been carried out.
机译:考虑到它们在TOF PET模块和PET / MRI混合系统中的可能应用,这里讨论了有关意法半导体最新的硅光电倍增管(SiPM)器件的时序和能量分辨率特性研究的初步结果。一对具有60×60微电池和3.5×3.5 mm〜2的有效面积的器件已与3×3×15 mm〜3 LYSO晶体耦合,并已充分表征。已经测量了来自不同γ源的线性响应以及来自〜(22)Na正电子源的两个511 keV光子的定时分辨率。