We have developed a front-end ASIC (MPPC32) intended for future PET scanners that offers time-of-flight (TOF) capability in conjunction with a multi-pixel photon counter (MPPC) array. The ASIC design is based on the open-IP project proposed by JAXA and was realized in TSMC 0.35-μm CMOS technology. The circuit comprises 32-channel, low impedance CMOS current conveyors (CCs) to effectively acquire fast MPPC signals. In order to precisely measure the coincidence timing of 511 keV gamma rays, the leading-edge method was employed instead of conventional zero-crossing measurement to discriminate signals. As a result, we obtained time jitter and walk measurement of 67 ps (FWHM) and 98 ps (within 511 keV ± 20%), respectively. Moreover, excellent energy resolutions of 9.8% (662 keV; FWHM) and 10.5% (511 keV; FWHM) were obtained by utilizing a 3 × 3 mm~2 MPPC (of 50 μm pitch) coupled with a Ce-doped LYSO (Ce:LYSO) crystal 3 × 3 × 10 mm~3 in size. We finally report on the TOF measurements, and demonstrate that the MPPC32 developed here can be a promising device for future TOF-PET scanners using the MPPC array.
机译:我们已经开发了一种面向未来的PET扫描仪的前端ASIC(MPPC32),它具有飞行时间(TOF)功能以及多像素光子计数器(MPPC)阵列。 ASIC设计基于JAXA提出的open-IP项目,并以TSMC0.35-μmCMOS技术实现。该电路包括32通道,低阻抗CMOS电流传输器(CC),可有效采集快速MPPC信号。为了精确地测量511 keV伽马射线的重合时间,采用前沿方法代替常规的过零测量来区分信号。结果,我们分别获得了67 ps(FWHM)和98 ps(在511 keV±20%范围内)的时间抖动和步行测量。此外,通过使用3×3 mm〜2 MPPC(间距为50μm)与掺Ce的LYSO(Ce)结合,可获得9.8%(662 keV; FWHM)和10.5%(511 keV; FWHM)的出色能量分辨率:LYSO)晶体3×3×10 mm〜3。我们最终报告了TOF测量值,并证明了此处开发的MPPC32对于使用MPPC阵列的未来TOF-PET扫描仪而言可能是很有前途的设备。

著录项

  • 来源
  • 作者单位

    Faculty of Advanced Science and Engineering, Waseda University, 3-4-1 Okubo, Shinjuku, Tokyo, Japan;

    Faculty of Advanced Science and Engineering, Waseda University, 3-4-1 Okubo, Shinjuku, Tokyo, Japan;

    Institute of Space and Astronautical Science, JAXA, Sagamihara, Kanagawa, Japan;

    Faculty of Advanced Science and Engineering, Waseda University, 3-4-1 Okubo, Shinjuku, Tokyo, Japan;

    Faculty of Advanced Science and Engineering, Waseda University, 3-4-1 Okubo, Shinjuku, Tokyo, Japan;

    Solid State Division, Hamamatsu Photonics K.K., 1126-1 Ichino-cho, Hamamatsu, Shizuoka, Japan;

    Solid State Division, Hamamatsu Photonics K.K., 1126-1 Ichino-cho, Hamamatsu, Shizuoka, Japan;

    Solid State Division, Hamamatsu Photonics K.K., 1126-1 Ichino-cho, Hamamatsu, Shizuoka, Japan;

    Solid State Division, Hamamatsu Photonics K.K., 1126-1 Ichino-cho, Hamamatsu, Shizuoka, Japan;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    positron emission tomography; multi-pixel photon counter; gamma-rays; ASIC; open-IP;