An industrial type 2 MeV/20 kW ILU-6 electron beam accelerator is operational in India for development of applications and technology demonstration to the Indian Industry in the field of polymer modifications (cross-linking and degradation). In order to adjust the treatment conditions and to control the good operation of the accelerator, it is necessary to study dose response as a function of various operational parameters. An experimental procedure for calibrating Cellulose Triacetate (CTA) film strip dosimeters in electron beam using total absorption graphite calorimeters is discussed and compared it with the results obtained from gamma calibration. Dosimetry data for process application, where the irradiation target is plane polymer sheet, have been obtained under various operational parameters.
机译:印度正在运行一种工业型2 MeV / 20 kW ILU-6电子束加速器,用于在聚合物改性(交联和降解)领域向印度工业开发应用和技术演示。为了调整治疗条件并控制加速器的良好运行,有必要研究剂量响应与各种运行参数的关系。讨论了使用总吸收石墨量热计在电子束中校准纤维素三乙酸酯(CTA)膜条剂量计的实验程序,并将其与从γ校准获得的结果进行了比较。在各种操作参数下,已获得用于辐照目标为平面聚合物片材的过程应用的剂量学数据。