An active beamstop for accurate measurement of high intensity X-ray beams

机译:主动式光束停止器,用于精确测量高强度X射线束

获取原文
获取原文并翻译 | 示例

摘要

In order to accurately measure the X-ray flux and determine the dose absorbed by the sample, we have designed an active beamstop device. Based on the photoelectric effect, the device consists of a beamstop, an insulator, and a concentric photoelectron collector. Tests of this device on the macromolecular crystallography beamline (BL17U1) at the Shanghai Synchrotron Research Facility (SSRF) show that the active beamstop is sufficient for determining the X-ray flux with the relative error smaller than 0.2%. The device can be used as a non-invasive, real-time intensity monitor at the sample during data collection. It can also be used with standard collimated beams or slits and served as a warning when the beam alignment drifts.
机译:为了准确地测量X射线通量并确定样品吸收的剂量,我们设计了一个有源防束器。基于光电效应,该设备由一个束流阻挡器,一个绝缘体和一个同心光电子收集器组成。在上海同步加速器研究设施(SSRF)的大分子晶体学光束线(BL17U1)上对该设备进行的测试表明,主动光束停止器足以确定X射线通量,相对误差小于0.2%。该设备可以在数据收集过程中用作样品的无创实时强度监测仪。它也可以与标准准直光束或狭缝一起使用,并在光束对准发生偏移时作为警告。