PET/MRI insert using digital SiPMs: Investigation of MR-compatibility

机译:使用数字SiPM的PET / MRI插入物:MR兼容性研究

获取原文
获取原文并翻译 | 示例

摘要

In this work, we present an initial MR-compatibility study performed with the world's first preclinical PET/MR insert based on fully digital silicon photo multipliers (dSiPM). The PET insert allows simultaneous data acquisition of both imaging modalities and thus enables the true potential of hybrid PET/MRI. Since the PET insert has the potential to interfere with all of the MRI's subsystems (strong magnet, gradients system, radio frequency (RF) system) and vice versa, interference studies on both imaging systems are of great importance to ensure an undisturbed operation. As a starting point to understand the interference, we performed signal-to-noise ratio (SNR) measurements as well as dedicated noise scans on the MRI side to characterize the influence of the PET electronics on the MR receive chain. Furthermore, improvements of sub-components' shielding of the PET system are implemented and tested inside the MRI. To study the influence of the MRI on the PET performance, we conducted highly demanding stress tests with gradient and RF dominated MR sequences. These stress tests unveil a sensitivity of the PETs electronics to gradient switching.
机译:在这项工作中,我们提出了一个基于全球首个基于全数字硅光电倍增管(dSiPM)的临床前PET / MR插入物进行的MR兼容性研究。 PET插入件允许同时获取两种成像方式的数据,从而实现混合PET / MRI的真正潜力。由于PET插入物可能会干扰MRI的所有子系统(强磁体,梯度系统,射频(RF)系统),反之亦然,因此在两个成像系统上进行干扰研究对于确保不受干扰的操作至关重要。作为了解干扰的起点,我们在MRI端进行了信噪比(SNR)测量和专用噪声扫描,以表征PET电子设备对MR接收链的影响。此外,在MRI内部实施并测试了PET子组件的屏蔽性能。为了研究MRI对PET性能的影响,我们对梯度和RF为主的MR序列进行了苛刻的压力测试。这些压力测试揭示了PET电子器件对梯度转换的敏感性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号