A comparison of digital zero-crossing and charge-comparison methods for neutron/γ-ray discrimination with liquid scintillation detectors

机译:液体闪烁探测器对中子/γ射线鉴别的数字过零和电荷比较方法的比较

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, we have compared the performances of the digital zero-crossing and charge-comparison methods for n/γ discrimination with liquid scintillation detectors at low light outputs. The measurements were performed with a 2″ × 2″ cylindrical liquid scintillation detector of type BC501A whose outputs were sampled by means of a fast waveform digitizer with 10-bit resolution, 4 GS/s sampling rate and one volt input range. Different light output ranges were measured by operating the photomultiplier tube at different voltages and a new recursive algorithm was developed to implement the digital zero-crossing method. The results of our study demonstrate the superior performance of the digital zero-crossing method at low light outputs when a large dynamic range is measured. However, when the input range of the digitizer is used to measure a narrow range of light outputs, the charge-comparison method slightly outperforms the zero-crossing method. The results are discussed in regard to the effects of the quantization noise and the noise filtration performance of the zero-crossing filter.
机译:在本文中,我们比较了在低光输出下使用液体闪烁检测器对数字过零和电荷比较方法进行n /γ判别的性能。使用BC501A型2“×2”圆柱形液体闪烁探测器进行测量,其输出通过具有10位分辨率,4 GS / s采样率和1伏特输入范围的快速波形数字化仪采样。通过在不同的电压下操作光电倍增管来测量不同的光输出范围,并开发了一种新的递归算法来实现数字过零方法。我们的研究结果表明,当测量大动态范围时,数字过零方法在低光输出下具有出色的性能。但是,当使用数字化仪的输入范围来测量窄范围的光输出时,电荷比较方法的性能略优于零交叉方法。讨论了有关量化噪声和过零滤波器的噪声过滤性能的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号