Accurate dose delivery is crucial to the success of modern radiotherapy. To evaluate the dose actually delivered to patients, in-vivo dosimetry (IVD) is generally performed during radiotherapy to measure the entrance doses. In IVD, a build-up device should be placed on top of an in-vivo dosimeter to satisfy the electron equilibrium condition. However, a build-up device made of tissue-equivalent material or metal may perturb dose delivery to a patient, and requires an additional laborious and time-consuming process. We developed a novel IVD method using a look-up table of conversion ratios instead of a buildup device. We validated this method through a monte-carlo simulation and 31 clinical trials. The mean error of clinical IVD is 3.17% (standard deviation: 2.58%), which is comparable to that of conventional IVD methods. Moreover, the required time was greatly reduced so that the efficiency of IVD could be improved for both patients and therapists.
机译:准确的剂量输送对现代放射治疗的成功至关重要。为了评估实际交付给患者的剂量,通常在放射治疗期间进行体内剂量测定(IVD)以测量进入剂量。在IVD中,应将堆积装置放在体内剂量计的顶部,以满足电子平衡条件。然而,由与组织等效的材料或金属制成的堆积装置可能会扰乱向患者的剂量输送,并且需要额外的费力且费时的过程。我们开发了一种新颖的IVD方法,使用的是转换率查找表而不是构建设备。我们通过蒙特卡洛模拟和31个临床试验验证了该方法。临床IVD的平均误差为3.17%(标准偏差:2.58%),与常规IVD方法的误差相当。此外,大大减少了所需的时间,从而可以提高患者和治疗师的IVD效率。

著录项

  • 来源
  • 作者单位

    Department of Radiation Oncology, Pusan National University Yangsan Hospital, Yangsan, Republic of Korea,Medical Research Institute, Pusan National University Hospital, Busan, Republic of Korea;

    Department of Radiation Oncology, Pusan National University Yangsan Hospital, Yangsan, Republic of Korea;

    Department of Radiation Oncology, Pusan National University Yangsan Hospital, Yangsan, Republic of Korea;

    Department of Radiation Oncology, Pusan National University Hospital, Busan, Republic of Korea;

    Department of Radiological Science, Dongseo University, Busan, Republic of Korea;

    Department of Radiation Oncology, Pusan National University Hospital, Busan, Republic of Korea;

    Department of Radiation Oncology, Pusan National University Hospital, Busan, Republic of Korea;

    Department of Radiation Oncology, Pusan National University Hospital, Busan, Republic of Korea;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Radiotherapy; In-vivo dosimetry; Build-up device; Radiotherapy; Conversion ratio; Glass rod dosimeter;

    机译:放射治疗;体内剂量测定;集结装置;放射治疗;转换率;玻璃棒剂量计;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号