New method for evaluating effective recovery time and single photoelectron response in silicon photomultipliers

机译:评估硅光电倍增管有效恢复时间和单光电子响应的新方法

获取原文
获取原文并翻译 | 示例

摘要

The linearity of a silicon photomultiplier (SiPM) response depends on the number of APD cells and its effective recovery time and it is related to the intensity and duration of the detected light pulses. The aim of this study was to determine the effective recovery time on the basis of the measured SiPM response to light pulses of different durations. A closer analysis of the SiPM response to the light pulses shorter than the effective recovery time of APD cells led to a method for the evaluation of the single photoelectron response of the devices where the single photoelectron peak cannot be clearly measured. This is necessary in the evaluation of the number of fired APD cells (or the number of photoelectrons) in measurements with light pulses of various durations. Measurements were done with SiPMs manufactured by two companies: Hamamatsu and SensL.
机译:硅光电倍增管(SiPM)响应的线性度取决于APD电池的数量及其有效恢复时间,并且与检测到的光脉冲的强度和持续时间有关。这项研究的目的是根据所测量的SiPM对不同持续时间的光脉冲的响应来确定有效恢复时间。对短于APD电池有效恢复时间的光脉冲的SiPM响应的更仔细的分析导致了一种无法评估单个光电子峰的器件的单光电子响应评估方法。这在评估使用各种持续时间的光脉冲进行测量时,已发射的APD电池的数量(或光电子的数量)非常必要。使用Hamamatsu和SensL这两家公司生产的SiPM进行了测量。