首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >Calibration of a new experimental chamber for PIXE analysis at the Accelerator Facilities Division of Atomic Energy Centre Dhaka (AECD)
【24h】

Calibration of a new experimental chamber for PIXE analysis at the Accelerator Facilities Division of Atomic Energy Centre Dhaka (AECD)

机译:在达卡原子能中心(AECD)加速器设施处对用于PIXE分析的新实验室进行校准

获取原文
获取原文并翻译 | 示例

摘要

A new experimental chamber has been installed at the 3 MV Van de Graaff Accelerator Facilities Division in the Atomic Energy Centre, Dhaka, to perform different Ion Beam Analysis (IBA) techniques. The calibration of this new setup for Particle Induced X-ray Emission (PIXE) technique has been done using a set of thin MicroMatter standards and GUPIX (PIXE spectrum analysis software), which is explicated in this paper. The effective thicknesses of the beryllium window of the X-ray detector and of the different absorbers used were determined. For standardization, the so called instrumental constant H (product of detector solid angle and the correction factor for the setup) as function of X-ray energy were determined and stored inside the GUPIX library for further PIXE analysis.
机译:在达卡原子能中心的3 MV Van de Graaff加速器设施部门安装了一个新的实验室,以执行不同的离子束分析(IBA)技术。使用一组薄的MicroMatter标准和GUPIX(PIXE频谱分析软件)对这种用于粒子诱导X射线发射(PIXE)技术的新设置进行了校准,本文对此进行了阐述。确定了X射线探测器的铍窗和所使用的不同吸收体的有效厚度。为了进行标准化,确定了作为X射线能量函数的所谓仪器常数H(检测器立体角与设置的校正因子的乘积),并将其存储在GUPIX库中以进行进一步的PIXE分析。