The Korea Heavy Ion Medical Accelerator (KHIMA) has developed a superconducting cyclotron for the carbon therapy, which is called KIRAMS-430. The cyclotron is designed to accelerate only ~(12)C~(6+) ions up to the energy of 430 MeV/u. It uses two normal conducting RF cavities. The RF frequency is about 70.76 MHz. The nominal dee voltage is 70 kV at the center and 160 kV at the extraction. The RF cavity was designed with 4 stems by using CST microwave studio (MWS). In this paper, we represent the simulation results and the optimized design of the RF cavity for the KIRAMS-430.
机译:韩国重离子医学加速器(KHIMA)开发了一种用于碳疗法的超导回旋加速器,称为KIRAMS-430。回旋加速器设计为仅加速〜(12)C〜(6+)离子,直至能量达到430 MeV / u。它使用两个正常的传导RF腔。射频频率约为70.76 MHz。中心处的标称Dee电压为70 kV,提取处为160 kV。通过使用CST微波工作室(MWS)将RF腔设计为具有4个茎。在本文中,我们代表了KIRAMS-430的仿真结果和RF腔的优化设计。