首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >Calibration of a single hexagonal NaI(Tl) detector using a new numerical method based on the efficiency transfer method
【24h】

Calibration of a single hexagonal NaI(Tl) detector using a new numerical method based on the efficiency transfer method

机译:使用基于效率传递方法的新数值方法校准单个六角形NaI(Tl)检测器

获取原文
获取原文并翻译 | 示例

摘要

Gamma-ray detector systems are important instruments in a broad range of science and new setup are continually developing. The most recent step in the evolution of detectors for nuclear spectroscopy is the construction of large arrays of detectors of different forms (for example, conical, pentagonal, hexagonal, etc.) and sizes, where the performance and the efficiency can be increased. In this work, a new direct numerical method (NAM), in an integral form and based on the efficiency transfer (ET) method, is used to calculate the full-energy peak efficiency of a single hexagonal NaI(Tl) detector. The algorithms and the calculations of the effective solid angle ratios for a point (isotropic irradiating) gamma-source situated coaxially at different distances from the detector front-end surface, taking into account the attenuation of the gamma-rays in the detector's material, end-cap and the other materials in-between the gamma-source and the detector, are considered as the core of this (ET) method. The calculated full-energy peak efficiency values by the (NAM) are found to be in a good agreement with the measured experimental data.
机译:伽马射线探测器系统是广泛科学领域中的重要仪器,并且新的设置正在不断发展。核光谱检测器发展的最新步骤是构造各种形式(例如,锥形,五边形,六边形等)和尺寸的检测器大阵列,可以提高性能和效率。在这项工作中,以积分形式并基于效率转移(ET)方法的新的直接数值方法(NAM)用于计算单个六角形NaI(Tl)检测器的全能峰值效率。考虑到探测器材料,端部中的伽玛射线的衰减,同轴(距各向同性的辐射)伽玛源与探测器前端表面距离不同的点的有效立体角比率的算法和计算-cap和伽马射线源与检测器之间的其他材料被视为此(ET)方法的核心。发现(NAM)计算出的全能峰值效率值与测量的实验数据非常吻合。

著录项

  • 来源
  • 作者单位

    Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria, Egypt;

    Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria, Egypt;

    Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna, Russia,Institute for Nuclear Research and Nuclear Energy, Bulgarian Academy of Sciences, 1784 Sofia, Bulgaria;

    Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria, Egypt;

    Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna, Russia,Institute for Nuclear Research and Nuclear Energy, Bulgarian Academy of Sciences, 1784 Sofia, Bulgaria;

    Department of Medical Equipment Technology, Faculty of Allied Medical Sciences, Pharos University in Alexandria, Egypt;

    Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna, Russia;

    Physics Department, Faculty of Science, Alexandria University, 21511 Alexandria, Egypt;

    Frank Laboratory of Neutron Physics, Joint Institute for Nuclear Research, 141980 Dubna, Russia;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Hexagonal NaI(Tl) detector; Effective solid angle ratio; Full-energy peak efficiency; Numerical analytical method (NAM); Efficiency transfer method (ET);

    机译:六角NaI(Tl)检测器;有效立体角比;全能峰值效率;数值分析方法(NAM);效率转移法;