Preliminary results of a prototype C-shaped PET designed for an in-beam PET system

机译:为光束内PET系统设计的原型C形PET的初步结果

获取原文
获取原文并翻译 | 示例

摘要

Positron emission tomography (PET) can be utilized in particle beam therapy to verify the dose distribution of the target volume as well as the accuracy of the treatment We present an in-beam PET scanner that can be integrated into a particle beam therapy system. The proposed PET scanner consisted of 14 detector modules arranged in a C-shape to avoid blockage of the particle beam line by the detector modules. Each detector module was composed of a 9×9 array of 4.0 mm × 4.0 mm × 20.0 mm LYSO crystals optically coupled to four 29-mm-diameter PMTs using the photomultiplier-quadrant-sharing (PQS) technique. In this study, a Ceant4 Application for Tomographic Emission (GATE) simulation study was conducted to design a C-shaped PET scanner and then experimental evaluation of the proposed design was performed. The spatial resolution and sensitivity were measured according to NEMA NU2-2007 standards and were 6.1 mm and 5.61 cps/kBq, respectively, which is in good agreement with our simulation, with an error rate of 12.0%. Taken together, our results demonstrate the feasibility of the proposed C-shaped in-beam PET system, which we expect will be useful for measuring dose distribution in particle therapy.
机译:正电子发射断层扫描(PET)可以用于粒子束治疗中,以验证目标体积的剂量分布以及治疗的准确性。我们提供了一种可以集成到粒子束治疗系统中的束内PET扫描仪。建议的PET扫描仪由14个C形排列的检测器模块组成,以避免粒子束线被检测器模块阻塞。每个检测器模块由一个9×9阵列组成,该阵列由一个4.0 mm×4.0 mm×20.0 mm的LYSO晶体组成,这些晶体通过光电倍增象限共享(PQS)技术与四个直径为29 mm的PMT光耦合。在这项研究中,进行了Ceant4断层扫描应用(GATE)仿真研究,以设计C形PET扫描仪,然后对该提议的设计进行了实验评估。空间分辨率和灵敏度根据NEMA NU2-2007标准进行测量,分别为6.1 mm和5.61 cps / kBq,与我们的模拟结果非常吻合,错误率为12.0%。综上所述,我们的结果证明了拟议中的C形束内PET系统的可行性,我们希望该系统可用于测量颗粒疗法中的剂量分布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号