Magnetic proton recoil (MPR) neutron spectrometer is considered as a powerful instrument to measure deuterium-tritium (DT) neutron spectrum, as it is currently used in inertial confinement fusion facilities and large Tokamak devices. The energy resolution (ER) and neutron detection efficiency (NDE) are the two most important parameters to characterize a neutron spectrometer. In this work, the ER calibration for the MPR spectrometer was performed by using the HI-13 tandem accelerator at China Institute of Atomic Energy (CIAE), and the NDE calibration was performed by using the neutron generator at CIAE. The specific calibration techniques used in this work and the associated accuracies were discussed in details in this paper. The calibration results were presented along with Monte Carlo simulation results.
机译:质子反冲(MPR)磁中子能谱仪被认为是测量氘-(DT)中子能谱的有力工具,因为它目前在惯性约束聚变设施和大型Tokamak装置中使用。能量分辨率(ER)和中子探测效率(NDE)是表征中子能谱仪的两个最重要的参数。在这项工作中,使用中国原子能科学研究院(CIAE)的HI-13串联加速器对MPR光谱仪进行ER校准,并使用CIAE的中子发生器进行NDE校准。本文详细讨论了这项工作中使用的特定校准技术以及相关的精度。给出了校准结果以及蒙特卡洛模拟结果。