Development of a multi-element microdosimetric detector based on a thick gas electron multiplier

机译:基于厚气体电子倍增器的多元素微剂量检测仪的研制

获取原文
获取原文并翻译 | 示例
           

摘要

A prototype multi-element gaseous microdosimetric detector was developed using the Thick Gas Electron Multiplier (THGEM) technique. The detector aims at measuring neutron and gamma-ray dose rates for weak neutron-gamma radiation fields. The multi-element design was employed to increase the neutron detection efficiency. The prototype THGEM multi-element detector consists of three layers of tissue equivalent plastic hexagons and each layer houses a hexagonal array of seven cylindrical gas cavity elements with equal heights and diameters of 17 mm. The final detector structure incorporates 21 gaseous volumes. Owing to the absence of wire electrodes, the THGEM multi-element detector offers flexible and convenient fabrication. The detector responses to neutron and gamma-ray were investigated using the McMaster Tandetron ~7Li(p,n) neutron source. The dosimetric performance of the detector is presented in contrast to the response of a commercial tissue equivalent proportional counter. Compared to the standard TEPC response, the detector gave a consistent microdosimetric response with an average discrepancy of 8 % in measured neutron absorbed dose. An improvement of a factor of 3.0 in neutron detection efficiency has been accomplished with only a small degradation in energy resolution. However, its low energy cut off is about 6 keV/μm, which is not sufficient to measure the gamma-ray dose. This problem will be addressed by increasing the electron multiplication gain using double THGEM layers.
机译:使用厚气体电子倍增器(THGEM)技术开发了原型多元素气态微剂量检测器。该探测器旨在测量弱中子-伽马辐射场的中子和伽马射线剂量率。采用多元素设计来提高中子检测效率。原型THGEM多元素检测器由三层组织等效塑料六边形组成,每层均包含七个圆柱状气体腔体元素的六边形阵列,高度和直径相等,均为17毫米。最终的检测器结构包含21种气体体积。由于没有线电极,THGEM多元素检测器提供了灵活方便的制造方法。使用McMaster Tandetron〜7Li(p,n)中子源研究了探测器对中子和伽马射线的响应。与商用组织等效比例计数器的响应相比,提供了检测器的剂量性能。与标准TEPC响应相比,该检测器给出了一致的微剂量响应,中子吸收剂量的平均差异为8%。中子检测效率的系数提高了3.0,而能量分辨率仅有很小的下降。但是,它的低能量截止约为6 keV /μm,不足以测量伽玛射线剂量。这个问题将通过使用双THGEM层增加电子倍增增益来解决。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号