A Monte Carlo method based on the GEANT4 toolkit has been developed to correct the full-energy peak (FEP) efficiencies of a high purity germanium (HPGe) detector equipped with a low background shielding system, and moreover evaluated using summing peaks in a numerical way. It is found that the FEP efficiencies of (60)~Co, (133)~Ba and (152)~Eu can be improved up to 18% by taking the calculated true summing coincidence factors (TSCFs) correction into account. Counts of summing coincidence y peaks in the spectrum of (152)~Eu can be well reproduced using the corrected efficiency curve within an accuracy of 3%.
机译:已开发出基于GEANT4工具包的蒙特卡洛方法,以纠正配备有低背景屏蔽系统的高纯度锗(HPGe)检测器的全能峰(FEP)效率,并且使用数字求和峰进行评估。发现通过考虑计算出的真实总和重合因子(TSCF)校正,可以将(60)〜Co,(133)〜Ba和(152)〜Eu的FEP效率提高18%。使用校正后的效率曲线,可以在3%的精度内很好地重现(152)〜Eu光谱中的y峰值的总和计数。