The aim of this work is to examine, from a theoretical point of view, the full-energy peak determination by starting from total efficiency and peak-to-total calculations, which are widely used in gamma-ray spectrometry. The detectors here examined include five different sodium iodide NaI(Tl) detectors and three high purity germanium HPGe detectors. The full-energy peak efficiency was determined both analytically and numerically. In some cases, the discrepancy between the experimental values was less than 6%. Some differences between semi-empirical calculations, analytical calculations, and Monte Carlo methods are shown. In addition, some of these results were compared with available experimental data.
机译:这项工作的目的是从一个理论的观点来看,通过从总效率和峰 - 总计算开始,从完全效率和峰 - 总计算中进行全能峰值测定,这是广泛用于伽马射线光谱光谱法的。此处检测的探测器包括五种不同的碘化钠Nai(TL)探测器和三个高纯度锗HPGE探测器。全能峰值效率在分析和数值上确定。在某些情况下,实验值之间的差异小于6%。显示了半经验计算,分析计算和蒙特卡罗方法之间的一些差异。此外,将其中一些结果与可用的实验数据进行了比较。