【24h】

Performance recovery of long CsI(Tl) scintillator crystals with APD-based readout

机译:基于APD的读数可恢复长CsI(Tl)闪烁体晶体的性能

获取原文
获取原文并翻译 | 示例

摘要

CALIFA is the high efficiency and energy resolution calorimeter for the R~3B experiment at FAIR, intended for detecting high energy light charged particles and gamma rays in scattering experiments, and is being commissioned during the Phase-0 experiments at FAIR, between 2018 and 2020. It surrounds the reaction target in a segmented configuration with 2432 detection units made of long CsI(Tl) finger-shaped scintillator crystals. CALIFA has a 10 year intended operational lifetime as the R~3B calorimeter, necessitating measures to be taken to ensure enduring performance. In this paper we present a systematic study of two groups of 6 different detection units of the CALIFA detector after more than four years of operation. The energy resolution and light output yield are evaluated under different conditions. Tests cover the aging of the first detector units assembled and investigates recovery procedures for degraded detection units. A possible reason for the observed degradation is given, pointing to the crystal-APD coupling.
机译:CALIFA是FAIR的R〜3B实验的高效和能量分辨率量热仪,旨在检测散射实验中的高能带电粒子和伽马射线,并将于2018年至2020年在FAIR的0期实验期间进行调试它以2432个由长CsI(Tl)手指形闪烁体晶体制成的检测单元以分段配置围绕反应目标。作为R〜3B量热仪,CALIFA的预期使用寿命为10年,因此必须采取措施以确保持久的性能。在本文中,我们对经过四年以上运行的CALIFA检测器的两组6个不同检测单元进行了系统的研究。在不同条件下评估能量分辨率和光输出产量。测试涵盖组装的第一个探测器单元的老化情况,并研究退化的探测器单元的恢复程序。给出了观察到的退化的可能原因,指出了晶体-APD耦合。