A liquid scintillation detector for radioassay of gadolinium-loaded liquid scintillator for the LZ Outer Detector

机译:用于LZ外部探测器的载有liquid的液体闪烁器放射分析的液体闪烁探测器

获取原文
获取原文并翻译 | 示例

摘要

We report on the design and performance of the LUX-ZEPLIN (LZ) "Screener", a small liquid scintillator detector consisting of approximate to 23 kg of LAB-based gadolinium-loaded liquid scintillator (GdLS) to be used in the LZ Outer Detector. The Outer Detector will be filled with 17.3 tonnes of GdLS and will surround the central liquid xenon time projection chamber of LZ. Its primary function will be to tag neutron events in the liquid xenon which could mimic a WIMP dark matter signal. To meet the deadtime requirements for the Outer Detector, the radioimpurity levels in the GdLS must be kept below less than or similar to 0.07 mBq/kg. This background level corresponds to a rate of approximate to 50 Hz above an energy threshold of 100 keV.The Screener was operated in the ultra-low-background environment of the former LUX water shield in the Davis Laboratory at the Sanford Underground Research Facility for radioassay of the GdLS. Careful selection of detector materials and use of ultra-low-background PMTs allows the measurement of a variety of radioimpurities. The C-14/C-12 ratio in the scintillator is measured to be (2.83 +/- 0.06(stat.)+/- 0.01(sys.)) x 10(-1)7. Use of pulse shape discrimination allows the concentration of isotopes throughout the U-238, U-235, and Th-232 chains to be measured by fitting the collected spectra from alpha and beta events. We find that equilibrium is broken in the U-238 and Th-232 chains and that a significant portion of the contamination in the GdLS results from decays in the Ac-227 subchain of the U-235 series.Predictions for the singles rate in the Outer Detector are presented. The rate from radioimpurities above 100 keV in the GdLS is estimated to be 97.9 +/- 6.4 Hz, with 65.5 +/- 1.9 Hz resulting from alpha-decays.
机译:我们报告了LUX-ZEPLIN(LZ)“ Screener ”的设计和性能,LUX-ZEPLIN(Screen)是一种小型液体闪烁探测器,由大约23公斤基于LAB的g基液体闪烁器(GdLS)组成,用于LZ外探测器。外部探测器将装满17.3吨的GdLS,并将包围LZ的中央液态氙时间投射室。它的主要功能是标记液态氙中的中子事件,这可能模仿WIMP暗物质信号。为了满足外部检测器的空载时间要求,GdLS中的放射性杂质水平必须保持低于或小于0.07 mBq / kg。该背景水平对应于高于100 keV能量阈值的大约50 Hz的速率。筛选器在桑福德大学地下研究设施的戴维斯实验室中,前LUX水屏蔽的超低背景环境下进行放射分析GdLS。仔细选择检测器材料并使用超低背景PMT可以测量各种放射性杂质。闪烁体中的C-14 / C-12比率经测量为(2.83 +/- 0.06(stat。)+/- 0.01(sys。))x 10(-1)7。通过使用脉冲形状判别,可以通过拟合从α和β事件收集的光谱来测量整个U-238,U-235和Th-232链中同位素的浓度。我们发现U-238和Th-232链中的平衡被打破,并且GdLS中很大一部分污染是由U-235系列的Ac-227子链的衰变导致的。显示了外部检测器。 GdLS中来自高于100 keV的放射性杂质的速率估计为97.9 +/- 6.4 Hz,其中65.5 +/- 1.9 Hz是由α衰减引起的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号