The response of a position-sensitive Li-glass scintillator detector to alpha-particles from a collimated Am-241 source scanned across the face of the detector has been measured. Scintillation light was read out by an 8 x 8 pixel multi-anode photomultiplier and the signal amplitude for each pixel has been recorded for every position on a scan. The pixel signal is strongly dependent on position and in general several pixels will register a signal (a hit) above a given threshold. The effect of this threshold on hit multiplicity is studied, with a view to optimize the single-hit efficiency of the detector.
机译:已经测量了位置敏感的锂玻璃闪烁体探测器对扫描到整个探测器表面的来自准直Am-241光源的α粒子的响应。闪烁光由8 x 8像素的多阳极光电倍增管读出,并且每个像素的信号幅度已在扫描的每个位置记录下来。像素信号强烈依赖于位置,通常,几个像素将记录高于给定阈值的信号(命中)。研究了该阈值对命中多重性的影响,以优化检测器的单命中效率。