Detector block performance based on a monolithic LYSO crystal using a novel signal multiplexing method

机译:基于新型信号多路复用方法的单片LYSO晶体的检测器块性能

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Organ dedicated PET devices provide improved imaging performance when compared to whole body systems. The present study summarizes the test carried out to study a new detector block designed for an organ dedicated PET system. This block includes three novel components namely the scintillator geometry and a retroreflector layer coupled to the entrance face, the photosensor and the readout electronics. We used arrays of 12 x 12 SiPM photosensors with 3 x 3 mm(2) active area each and a pitch of 4.2 mm. We are proposing a new readout electronics that permits to reduce the 12 row and columns signals to only 8 without significant detector performance degradation. This approach also allows for resolving radioactive sources in the whole volume of the proposed crystal, significantly reducing the edge effect that typically rejects these events.An overall spatial resolution of about 1.8 mm FWHM is obtained for the whole scintillation volume, with an average energy resolution of 13% FWHM and a photon depth of interaction resolution (FWHM) of 3.7 mm. (C) 2018 Elsevier B.V. All rights reserved.
机译:与全身系统相比,器官专用的PET设备可提供更好的成像性能。本研究总结了为研究用于器官专用PET系统的新型检测器模块而进行的测试。该模块包括三个新颖的组件,即闪烁体几何形状和耦合到入射面,光电传感器和读出电子设备的后向反射器层。我们使用了12 x 12 SiPM光电传感器阵列,每个阵列的有效面积为3 x 3 mm(2),间距为4.2 mm。我们正在提议一种新的读出电子设备,它可以将12行和列信号减少到仅8个,而不会显着降低检测器性能。这种方法还可以解决所提议晶体的整个体积中的放射源,从而显着降低通常会拒绝这些事件的边缘效应。整个闪烁体的整体空间分辨率约为1.8 mm FWHM,平均能量分辨率FWHM为13%,相互作用的光子深度分辨率(FWHM)为3.7 mm。 (C)2018 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号