首页> 外文期刊>NEC Research & Development >Evaluation of a Multi-Channel EEG (Electroencephalogram) Measurement Electrode System
【24h】

Evaluation of a Multi-Channel EEG (Electroencephalogram) Measurement Electrode System

机译:多通道脑电图测量电极系统的评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This paper reports on the evaluation of the 42 ch ECI electro-cap system. During P300 experiments, the amount of time the cap was worn and each impedance were measured. After the experiment, participants were interviewed about the cap. The results showed that using this cap enabled us to fix all electrodes to a scalp within 30 minutes, with low impedance. The subjective evaluation results showed that in the two-hour-EEG (Electroencephalogram) measurement most participants felt little pain. This result proves that electro-cap system is useful of multi-channel EEG measurements not only for brain research but also for clinical testing.