首页> 外文期刊>Microelectromechanical Systems, Journal of >A SU-8-Based Fully Integrated Biocompatible Inductively Powered Wireless Neurostimulator
【24h】

A SU-8-Based Fully Integrated Biocompatible Inductively Powered Wireless Neurostimulator

机译:基于SU-8的完全集成生物相容性感应供电的无线神经刺激仪

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A fully integrated minimally invasive compact polymer-based wireless neurostimulator was designed, fabricated, and characterized both in open air and in vivo. The neurostimulator $(3.1 times 1.6 times 0.3 hbox{mm})$ consists of a planar spiral coil for wireless power supply through inductive coupling, two Schottky diodes for full-wave rectification, an application-specific integrated circuit neurostimulator circuit chip for stimulus spike signal generation, and two biphasic platinum–iridium (PtIr) stimulation electrodes. The device is fully integrated and completely embedded in biocompatible SU-8 packaging. For in vivo testing, the wireless neurostimulator was implanted subcutaneously in a rat hind limb. At the coupling power of 21 dBm (125 mW) at 394-MHz resonant frequency, stable and robust cortical responses during extended periods of wireless stimulation were recorded.$hfill$[2012-0085]
机译:在露天和体内均设计,制造并表征了完全集成的微创紧凑型基于聚合物的无线神经刺激器。神经刺激器$(3.1乘以1.6乘以0.3 hbox {mm})$由用于感应电源的无线供电的平面螺旋线圈,用于全波整流的两个肖特基二极管,用于刺激尖峰的专用集成电路神经刺激器电路芯片组成信号生成和两个双相铂铱铱(PtIr)刺激电极。该设备完全集成并完全嵌入生物兼容的SU-8包装中。为了进行体内测试,将无线神经刺激器皮下植入大鼠后肢。在394-MHz共振频率下的21 dBm(125 mW)耦合功率下,记录了无线刺激长时间内稳定且鲁棒的皮层响应。$ hfill $ [2012-0085]