首页> 外文期刊>Measurement Techniques >INTERNATIONAL COLLABORATION: INTERNATIONAL COMPARISON OF THE SPECTRAL RESPONSIVITIES OF SILICON PHOTODETECTORS
【24h】

INTERNATIONAL COLLABORATION: INTERNATIONAL COMPARISON OF THE SPECTRAL RESPONSIVITIES OF SILICON PHOTODETECTORS

机译:国际合作:硅光电检测器光谱责任的国际比较

获取原文
获取原文并翻译 | 示例
           

摘要

The results are considered of a comparison of the national scales of spectral responsively of 18 national meteorological laboratories including the All-Union Scientific-Research Institute of Optophysical Measurements (VNIIOFI) between themselves and with the scale of the International Bureau of Weights and Measures (BIPM) in the spectral wavelength range 250-1000 nm using two different types of silicon photodetector. The comparisons indicated satisfactory agreement between th3 results of measurements of spectral responsively by the participating laboratories. The difference between the spectral responsively scale of VNIIOFI and that of the BIPM did not exceed ± 0.5/100 for the visible wavelength range and the near infrared. The maximum difference of the scales in the near ultraviolet wavelength range was 3.7/100.
机译:考虑将结果与18个国家气象实验室的光谱响应规模进行了比较,其中包括全联盟光物理测量科学研究所(VNIIOFI)与国际度量衡局(BIPM)的规模使用两种不同类型的硅光电探测器在250-1000 nm的光谱波长范围内)。比较表明参与实验室响应光谱的测量结果之间令人满意的一致性。对于可见光波长范围和近红外光谱,VNIIOFI和BIPM的光谱响应标度之间的差异不超过±0.5 / 100。在近紫外波长范围内水垢的最大差为3.7 / 100。

著录项

  • 来源
    《Measurement Techniques》 |1999年第1期|86-94|共9页
  • 作者

    R.I. Stoyarevskaya;

  • 作者单位
  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);