首页> 外文期刊>Journal of testing and evaluation >Chloride-Induced Stress Corrosion Cracking Tester for Austenitic Stainless Steel
【24h】

Chloride-Induced Stress Corrosion Cracking Tester for Austenitic Stainless Steel

机译:奥氏体不锈钢的氯化物诱导应力腐蚀裂解测试仪

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, a new experimental tester was designed to improve the stability in load measurementand imposed environment for chloride-induced stress corrosion cracking (CISCC)testing of austenitic stainless steels that are used to construct spent nuclear fuel dry storagecanisters. By fabricating compact tension specimens using the UNS S30400 stainless steel,the test is conducted by immersing the specimen in 5 % salinity artificial seawater at 50°C.From the measured load, the stress intensity factor is calculated and the crack length ismeasured by an optical microscope. The present experimental data confirm the existing findingsthat the CISCC growth rate is insensitive to the stress intensity factor but is dominantlyaffected by the specimen surface temperature. The crack growth rate also confirms theexisting model.
机译:在本文中,设计了一种新的实验测试仪以提高负载测量的稳定性氯化物诱导应力腐蚀裂纹(CISCC)强加的环境用于构建核燃料干储存的奥氏体不锈钢的测试罐。通过使用UNS S30400不锈钢制造紧凑的张力样品,通过将样品浸入5%盐度人工海水中在50℃下浸入试样进行测试。从测量的负载,计算应力强度因子,并且裂缝长度是通过光学显微镜测量。目前的实验数据证实了现有的发现CISCC增长率对压力强度因子不敏感,但主要是占主导地位的受试样的影响。裂缝增长率也证实了现有模型。