In this study, the repeatability (variation in measurements) and reproducibility (precision between measuring persons) of a measuring method, developed by us, for quantifying the visibility of guidewires was investigated. Visibility was determined on a percentage scale against the background. The measuring method was tested by two persons with three repeats using X-ray images. Images were made from ten guidewires with the same material and type. X-ray images were made with the same settings on different days. Coefficients of variation (standard deviation as a proportion of the mean) for three repeats did not exceed 0.10 in measurement. Therefore, our method was repeatable. The average difference between the persons was 4 %, therefore our method was reproducible. Coefficients of variation for the mean of the ten images were under 0.10. Therefore, the images made with the same settings on different days could be averaged in case of the same samples. Significantly, there was a difference (ρ = 0.04) between the three tests by the first person. There were no significant differences in case of the second person (ρ = 0.74), and between the three-three test (ρ = 0.21). Using this method, available guidewires can be tested under laboratory conditions, so they may become comparable based on their visibility.
机译:在这项研究中,我们研究了一种用于量化导丝可见度的测量方法的可重复性(测量中的差异)和可重复性(测量人员之间的精度)。可见度是在相对于背景的百分比范围内确定的。该测量方法由两个人使用X射线图像进行了三次重复测试。图像由十根相同材质和类型的导线制成。 X射线图像是在不同的日子以相同的设置制作的。测量的三个重复的变异系数(标准偏差作为平均值的比例)不超过0.10。因此,我们的方法是可重复的。人与人之间的平均差异为4%,因此我们的方法具有可重复性。十张图像的平均值的变异系数在0.10以下。因此,在相同样本的情况下,可以对在不同日期使用相同设置制作的图像进行平均。值得注意的是,第一人称的三项测试之间存在差异(ρ= 0.04)。在第二人称情况下(ρ= 0.74),在三者三项检验之间(ρ= 0.21)没有显着差异。使用这种方法,可以在实验室条件下对可用的导线进行测试,因此根据其可见度,它们可以变得可比。

著录项

  • 来源
    《Journal of testing and evaluation》 |2016年第6期|2085-2091|共7页
  • 作者

    Liza Pelyhe; Eszter Bognar;

  • 作者单位

    Faculty of Mechanical Engineering, Dept.of Materials Science and Engineering,Budapest Univ. of Technology and Economics, Bertalan Lajos Street 7, H-1111 Budapest,Hungary;

    Faculty of Mechanical Engineering, Dept. of Materials Science and Engineering,Budapest Univ. of Technology and Economics, Bertalan Lajos Street 7, H-1111 Budapest, Hungary,MTA-BME Research Group for Composite Science and Technology, Muegyetem rkp. 3., H-1111 Budapest,Hungary;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    guidewire; quantification; radiopacity; visibility;

    机译:导丝;量化射线不透性;能见度;