摘要

Many types of hip prosthesis exist on the market, and no standard exists for testing their functionality. There is a lack of agreement in the literature about the relevance of several design parameters on the load transfer between the femoral stem and bone. This can be explained considering the lack of investigations on the methodology of load transfer measurements. A testing protocol for such purpose must fulfil three requirements it must be physiological, it must be repeatable and reproducible and it must be sensitive to different prostheses. A methodological investigation is presented that was necessary to design a repeatable and reproducible protocol. The loading and constraints were derived from physiological data. The set-up and protocol were enhanced by means of (a) selection of an appropriate test specimen, (b) definition of reproducible reference systems, (c) sensitivity analyses on the sources of experimental error and (d) modification of the loading system to minimize the sources of error. The variability of results was as low as 0.8 per cent for repetitions on the same specimen, and 11 per cent between femurs. The capability of the protocol of detecting differences between stems was tested on different prostheses, including a prototype device. Significant differences between the load transfer patterns for the different designs were detected. The prototype device yielded satisfactory results.
机译:市场上存在许多类型的髋关节假体,并且没有用于测试其功能的标准。关于股骨干和骨之间负荷转移的几种设计参数的相关性,文献上缺乏共识。考虑到缺乏对负荷转移测量方法的研究,可以对此进行解释。为此目的的测试方案必须满足三个要求,必须是生理要求的,必须是可重复的和可再现的,并且必须对不同的假体敏感。提出了一种方法学调查,这对于设计可重复和可重现的协议是必需的。负荷和约束条件是根据生理数据得出的。通过(a)选择合适的试样,(b)定义可重复的参考系统,(c)对实验误差来源的敏感性分析和(d)修改加载系统来增强设置和协议尽量减少错误源。在同一标本上重复的结果变异性低至0.8%,而股骨之间的变异性低至11%。在包括原型设备在内的不同假体上测试了协议检测茎之间差异的能力。检测到了不同设计的载荷传递模式之间的显着差异。该原型装置产生了令人满意的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号