COMPARATIVE STUDY BETWEEN INSTRUMENTAL NEUTRON ACTIVATION ANALYSIS AND INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY IN ANALYSIS OF ENVIRONMENTAL SAMPLES

机译:仪器中子活化分析与电感耦合等离子体原子发射光谱法在环境样品分析中的比较研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Elemental analysis of different environmental Certified Reference Materials (CRM) was carried out using Instrumental Neutron Activation Analysis (INAA) and Inductively Coupled Plasma-Atomic Emission Spectrometry (ICP-AES) followed by Microwave digestion. The determination of about 30 elements in CRMs using INAA has been achieved with high accuracy and precision. The relative error of most of the elements is within 10% and the standard deviation is less than 15%. Z-score as an analytical performance is also less than +-2. ICP-AES was also found as a suitable method for analysis of soil and hair samples. However, INAA was found more suitable than ICP-AES for the determination of elements in filter samples. On the other hand, INAA appeared inadequate for the analysis of Ca because of the low count rate, Both methods can be proposed as complementary methods for elemental analysis. A combination of these two methods may offer the most valuable information on elemental analysis. Multiple analyses of the same element using various methods would ensure the quality of the data.
机译:使用仪器中子活化分析(INAA)和电感耦合等离子体原子发射光谱法(ICP-AES)对不同的环境认证参考材料(CRM)进行元素分析,然后进行微波消解。使用INAA可以确定CRM中CRM中约30种元素的准确性和准确性。大多数元素的相对误差在10%以内,标准偏差小于15%。 Z值作为分析性能也小于+ -2。还发现ICP-AES是分析土壤和头发样品的合适方法。但是,发现INAA比ICP-AES更适合测定过滤器样品中的元素。另一方面,由于计数率低,INAA似乎不足以分析Ca。这两种方法均可作为元素分析的补充方法。这两种方法的组合可以提供有关元素分析的最有价值的信息。使用各种方法对同一元素进行多次分析将确保数据的质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号