An ‘‘atom buncher’’ for controlling the concentration of gaseous samples has been conceptualized, evaluated theoretically, fabricated, and tested with excellent results. In effect, the atom buncher greatly increases the probability that a free atom will be in a small detector volume at a desired time. This was accomplished by using cryogenic techniques to condense atoms on a small spot and a pulsed laser to momentarily heat the spot to release the atoms at the desired time. Our work on noble gas atom counting by using resonance ionization spectroscopy is discussed as one example of the applications of the atom buncher.
机译:用于控制气态样品浓度的“原子聚束器”已被概念化,在理论上进行了评估,制造和测试,均获得了出色的结果。实际上,原子聚集器极大地增加了自由原子在所需时间处于较小检测器体积中的可能性。这是通过使用低温技术将原子凝聚在一个小点上并使用脉冲激光瞬时加热该点以在所需时间释放原子来实现的。我们讨论了使用共振电离光谱法对稀有气体原子计数的工作,作为原子聚集器应用的一个例子。