首页> 外文期刊>Journal of Applied Physics >Dead Time and Non‐Linearity Characteristics of the Geiger‐Counter X‐Ray Spectrometer
【24h】

Dead Time and Non‐Linearity Characteristics of the Geiger‐Counter X‐Ray Spectrometer

机译:盖革计数器X射线光谱仪的死区时间和非线性特性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The satisfactory use of Geiger‐counter methods for the measurement of diffracted x‐ray intensities demands that the observed counting rates be corrected for the non‐linear response of the counter. Both the multiple foil method of calibrating the non‐linearity of response and the two‐source method of Beers for measuring counter dead time have been found to be unsatisfactory approaches to this problem. The experience of this laboratory indicates that electronically controlled oscillographic techniques for the direct measurement of dead time are much more satisfactory. With a knowledge of the dead time, τ, the true counting rate, N, can be calculated from the observed counting rate, No, by employing the formula N=No/(1-Noτ). For a pulsating x‐ray source the effective counting rate is faster than the observed by a factor which is characteristic of the type of rectification employed, the target material, and the voltage applied to the x‐ray tube. A sufficiently accurate value of this factor can be arrived at by oscillographic observation of the pulse pattern and in some cases by mathematical analysis.
机译:Geiger计数器方法在衍射X射线强度测量中的令人满意的使用要求针对计数器的非线性响应对观察到的计数率进行校正。已经发现,用于校正响应非线性的多重箔法和用于测量反死时间的比尔斯的双源方法均不能令人满意地解决该问题。该实验室的经验表明,用于直接测量死区时间的电子控制示波器技术更加令人满意。知道死区时间τ后,可以通过使用公式N = No /(1-Noτ)从观察到的计数率No中计算出真实计数率N。对于脉动X射线源,有效计数率比所观察到的要快,这是所采用的整流类型,目标材料和施加在X射线管上的电压的特征。可以通过示波器观察脉搏图并在某些情况下通过数学分析得出该因子的足够准确的值。

著录项

  • 来源
    《Journal of Applied Physics》 |1949年第8期|共6页
  • 作者

  • 作者单位
  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号