A current-compensating technique for making a high-voltage current-comparator-based capacitance bridge insensitive to frequency fluctuations when used for measuring shunt reactor loss is described. A compensating current, proportional to twice the change in the current through the reference capacitor due to frequency fluctuations, is forced into the comparator winding through which the current from the reference capacitor passes resulting in a current in that winding having frequency characteristics similar to those of the reactor current. This compensating current is derived from a reduced replica of the applied high voltage or from the current through the reference capacitor. It need not be supplied with great accuracy for it is only a small fraction of the reference current. Thus the effect of frequency fluctuations on the inductive balance of the bridge is eliminated without sacrificing the inherent high accuracy of the high-voltage current-comparator-based capacitance bridge.
机译:描述了一种电流补偿技术,该电流补偿技术用于使高压基于电流比较器的电容电桥在测量并联电抗器损耗时对频率波动不敏感。与由于频率波动而流经参考电容器的电流变化成两倍的补偿电流被强制进入比较器绕组,来自参考电容器的电流流经比较器绕组,从而在该绕组中产生的电流具有与以下特性相似的频率特性:电抗器电流。该补偿电流源自施加的高电压的减小的副本或源自参考电容器的电流。由于它只是参考电流的一小部分,因此无需高精度地供电。因此,在不牺牲基于高压电流比较器的电容电桥固有的高精度的情况下,消除了频率波动对电桥电感平衡的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号