首页> 外文期刊>IEEE Transactions on Nuclear Science >A prototype BaF/sub 2//TMAE low pressure multiwire proportional chamber for PET
【24h】

A prototype BaF/sub 2//TMAE low pressure multiwire proportional chamber for PET

机译:用于PET的原型BaF / sub 2 // TMAE低压多线比例室

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A prototype annihilation photon detector combining a 50 mm*50 mm*8 mm BaF/sub 2/ scintillator with an MWPC (multiwire proportional chamber) filled with a photoionizing vapor (TMAE) at low pressure has been constructed. The structure of the detector is currently undergoing optimization but its present performance includes 8 mm FWHM (full width at half maximum) spatial resolution, 12-ns timing resolution, and 19% efficiency. It is intended that this technology, once proven, should form the basis for a clinical PET (positron emission tomography) system. In order to cover the large active areas (600 mm*400 mm) of a clinical system, it will be necessary to use tiles of BaF/sub 2/. Crystal boundary effects on image quality, studied using Monte Carlo methods, show degraded spatial resolution by 40% axially and 27% radially/tangentially for a 6*4 array of 16 mm-thick crystals replacing a hypothetical large single crystal. Further studies to characterize the distribution of scattered events in the proposed camera are described.
机译:构造了一个原型an灭光子检测器,该检测器将50 mm * 50 mm * 8 mm BaF / sub 2 /闪烁体与低压充满光电离蒸气(TMAE)的MWPC(多线比例室)组合在一起。检测器的结构目前正在优化中,但其当前性能包括8 mm FWHM(半高全宽)空间分辨率,12 ns定时分辨率和19%的效率。旨在证明该技术一旦获得验证,就应构成临床PET(正电子发射断层扫描)系统的基础。为了覆盖临床系统的较大活动区域(600毫米* 400毫米),必须使用BaF / sub 2 /的瓷砖。使用蒙特卡洛方法研究的晶体边界对图像质量的影响显示,对于6 * 4的16毫米厚晶体阵列,其轴向分辨率下降了40%,径向/切线径向分辨率降低了27%,从而取代了假设的大型单晶。描述了表征所提议照相机中的分散事件分布的进一步研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号