The Direct Detection Effect in the Hot-Electron Bolometer Mixer Sensitivity Calibration

机译:热电子鼓泡仪混合器灵敏度校准中的直接检测效果

获取原文
获取原文并翻译 | 示例

摘要

We investigate an error in the noise temperature measurements of the hot-electron bolometer mixers caused by the so-called "direct detection effect". The effect originates in the changing of the mixer parameters when the mixer is loaded on calibration black body sources at different temperatures (300 and 77 K). A correction factor was obtained from the mixer output power versus the bias current dependence, measured by: 1) the local oscillator (LO) power tuning: 2) mixer heating: and 3) application of an external RF source. Furthermore, the direct detection effect was assessed by elimination of the heterodyne response using a LO frequency, which is far off the mixer RF band. We show that the direct detection effect can be mitigated by using an isolator between the mixer and the IF amplifier
机译:我们调查了由所谓的“直接检测效应”引起的热电子辐射热计混频器噪声温度测量中的误差。当混频器在不同温度(300和77 K)下加载到校准黑体源上时,其影响源自混频器参数的变化。从混频器输出功率与偏置电流相关性获得了一个校正因子,该修正因子的测量方法如下:1)本地振荡器(LO)功率调整:2)混频器加热:和3)外部RF源的施加。此外,通过使用LO频率消除外差响应来评估直接检测效果,该LO频率远离混频器RF频段。我们表明,通过在混频器和IF放大器之间使用隔离器可以减轻直接检测效果