首页> 外文期刊>Biomedical Circuits and Systems, IEEE Transactions on >A 300-mV 220-nW Event-Driven ADC With Real-Time QRS Detection for Wearable ECG Sensors
【24h】

A 300-mV 220-nW Event-Driven ADC With Real-Time QRS Detection for Wearable ECG Sensors

机译:具有可穿戴式ECG传感器的实时QRS检测功能的300mV 220nW事件驱动ADC

获取原文
获取原文并翻译 | 示例
           

摘要

This paper presents an ultra-low-power event-driven analog-to-digital converter (ADC) with real-time QRS detection for wearable electrocardiogram (ECG) sensors in wireless body sensor network (WBSN) applications. Two QRS detection algorithms, pulse-triggered (PUT) and time-assisted PUT (t-PUT), are proposed based on the level-crossing events generated from the ADC. The PUT detector achieves 97.63% sensitivity and 97.33% positive prediction in simulation on the MIT-BIH Arrhythmia Database. The t-PUT improves the sensitivity and positive prediction to 97.76% and 98.59% respectively. Fabricated in 0.13 CMOS technology, the ADC with QRS detector consumes only 220 nW measured under 300 mV power supply, making it the first nanoWatt compact analog-to-information (A2I) converter with embedded QRS detector.
机译:本文介绍了一种用于无线人体传感器网络(WBSN)应用中可穿戴心电图(ECG)传感器的具有实时QRS检测的超低功耗事件驱动模数转换器(ADC)。基于ADC产生的电平交叉事件,提出了两种QRS检测算法:脉冲触发(PUT)和时间辅助PUT(t-PUT)。在MIT-BIH心律失常数据库的仿真中,PUT检测器可实现97.63%的灵敏度和97.33%的阳性预测。 t-PUT将灵敏度和阳性预测分别提高到97.76%和98.59%。具有QRS检测器的ADC采用0.13 CMOS技术制造,在300 mV电源下测得的功耗仅为220 nW,使其成为首款具有嵌入式QRS检测器的纳瓦紧凑型模数(A2I)转换器。