首页> 外文期刊>Geotechnical testing journal >Laboratory Calibration of Earth Pressure Cells
【24h】

Laboratory Calibration of Earth Pressure Cells

机译:土压室的实验室校准

获取原文
获取原文并翻译 | 示例
       

摘要

The objective of this study was to design a simple apparatus for laboratory calibration of diaphragm type earth pressure cells (EPC). A new testing device was fabricated to permit the application of normal stress to the EPC using various types of soil and load configurations. With respect to fluid pressure, the response of the EPC can be different for the simplest earth pressure conditions because of an arching effect or nonuniform contact stress. For an EPC with a rigid outer rim, sensitivities computed from soil calibrations were lower than those determined from fluid calibrations by about 20 %.
机译:这项研究的目的是设计一种简单的仪器,用于实验室校准隔膜型土压传感器(EPC)。制造了一种新的测试设备,以允许使用各种类型的土壤和负载配置将正应力施加到EPC上。关于流体压力,由于拱起效应或不均匀的接触应力,对于最简单的土压力条件,EPC的响应可能会有所不同。对于具有刚性外缘的EPC,根据土壤校准计算出的灵敏度比根据流体校准确定的灵敏度低约20%。